DRX-ACV
AC VOLTAGE INPUT SIGNAL CONDITIONER
Made in the USA
INFINITY
Electromagentic Compatibility
Year 2000 Compliant
What You need:
ò Your choice of the signal conditioner
ò IBM PC or compatible
ò Windows 3.1 or higher, or Windows 95
ò RJ12 connector with telephone cable
ò Adapter (DB9/25 to RJ12)
SETTING UP YOUR SYSTEM
DRN Computer Connection
INPUT RANGES
400 mV, 4V, 40 Volt, 400 Volt

DRX-ACVFREQUENCY RANGE
30 Hz to 1KHz

INPUT IMPEDANCE

2.1 Meg for all ranges

ISOLATION

Three way isolation up to 1800 V Peak for 1 minute, 354 V Peak continuous per IEC spacing Common mode rejection 100 dB

INPUT OVER-CURRENT PROTECTION

  • 10% Above full scale continuously
  • 100% Above full scale for 10 seconds


  • ANALOG TO DIGITAL TECHNIQUE

    Dual slope

    READ RATE

    3 readings/second

    ACCURACY AT 25 ░C

  • 400 mV, 4V, 40V and 400V ranges
  • 49 Hz to 500 Hz ▒0.2% of FS
  • 30 Hz to 1KHz ▒0.2% of FS ▒10 counts


  • TEMPERATURE STABILITY

  • 400 mV and 40 Volt range 150 ppm/░C typical
  • 4V and 400 Volt range 100 ppm/░C typical


  • STEP RESPONSE FOR RS485 OUTPUT

    2 seconds to 99% of the final value (Filter time constant = 64)

    RESPONSE TIME

    To verify the response time, check the carriage return <CR>, it will be sent at the end of the response. You can send another command after you receive the <CR>.
    i.e. send: *01X01 / response: 01X01<DATA><CR>
    Note:
    01 is address.

    Signal Conditioner
    WARM UP TO RATED ACCURACY
    30 minutes

    INPUT POWER

    10 to 32 Volt

    DC POWER CONSUMPTION

    2.4 Watts (100 mA at 24 V DC)

    OPERATING AMBIENT

    -5 to +55 ░C

    STORAGE TEMPERATURE

    -40 to +85 ░C

    RELATIVE HUMIDITY

    90% at 40 ░C non-condensing

    Help Price List Specifications Print Manual Manuals Home
     Home Home  Table of Contents TOC  Back Back  Next Next